Analytical Instrumentation

Read revision 6

Revision 6 · Oct 12, 2026 +32 −5

Datasheet constraints: every range, checkbox and listbox field now states its direction
Showing changes from Rev 5 to Rev 6 in Analytical Instrumentation.
---
title: Analytical Instrumentation
…128 unchanged lines
- "Instrument index entries and loop diagrams"
- "Mounting and installation details"
+constraint: all_of
```
…22 unchanged lines
- "Parts and consumables availability confirmation"
- "Startup and verification plan"
+constraint: all_of
```
…35 unchanged lines
- "Warranty documentation"
- "Spare parts receipt"
+constraint: all_of
```
…5 unchanged lines
```datasheet
−label: Minimum Manufacturer Experience with the Analyzer Type
+label: Manufacturer Experience with the Analyzer Type
type: range
unit: years
…2 unchanged lines
max: 15
setpoints: [0, 3, 5, 10, 15]
+constraint: min
```
…93 unchanged lines
```datasheet
−label: Minimum Ambient Operating Temperature Rating
+label: Lowest Ambient Operating Temperature Rating
type: range
unit: "°C"
…4 unchanged lines
setpoints: [-40, -30, -20, -10, 0, 5]
default: derived
+constraint: max
```
### The maximum ambient operating temperature rating of analyzers and transmitters mounted outdoors or in an unconditioned space shall be as indicated in the datasheet.
```datasheet
−label: Maximum Ambient Operating Temperature Rating
+label: Highest Ambient Operating Temperature Rating
type: range
unit: "°C"
…4 unchanged lines
setpoints: [40, 45, 50, 55, 60]
default: derived
+constraint: min
```
…42 unchanged lines
- "Streaming current"
- "Dissolved ozone"
+constraint: all_of
```
…93 unchanged lines
max: 10000
setpoints: [1, 10, 100, 1000, 4000, 10000]
+constraint: exact
```
…32 unchanged lines
- "Total chlorine"
- "Monochloramine"
+constraint: all_of
```
…19 unchanged lines
max: 20
setpoints: [0.5, 1, 2, 5, 10, 20]
+constraint: exact
```
…31 unchanged lines
setpoints: [0.02, 0.2, 2, 20, 40, 60]
default: 20
+constraint: exact
```
…27 unchanged lines
max: 2000000
setpoints: [2, 20, 200, 2000, 20000, 200000, 2000000]
+constraint: exact
```
…10 unchanged lines
setpoints: [0.01, 0.05, 0.1, 0.5, 1, 10]
default: derived
+constraint: exact
```
…39 unchanged lines
max: 1000
setpoints: [1, 5, 20, 50, 100, 1000]
+constraint: exact
```
…23 unchanged lines
max: 1000
setpoints: [2, 5, 20, 50, 100, 1000]
+constraint: exact
```
…23 unchanged lines
max: 10000
setpoints: [0.5, 1, 5, 20, 50, 100, 1000, 10000]
+constraint: exact
```
…25 unchanged lines
max: 500000
setpoints: [1000, 5000, 30000, 100000, 300000, 500000]
+constraint: exact
```
…54 unchanged lines
max: 20
setpoints: [0.5, 2, 5, 20]
+constraint: exact
```
…13 unchanged lines
- "Submersion in an open basin or channel on a rail, float, or handrail mount"
- "Flow-through cell on a sample panel fed from a sample tap"
+constraint: all_of
```
…32 unchanged lines
```datasheet
−label: Maximum Sample Transport Lag
+label: Sample Transport Lag
type: range
unit: s
…2 unchanged lines
max: 600
setpoints: [10, 30, 60, 120, 300, 600]
+constraint: max
```
…27 unchanged lines
- "Low-sample-flow switch wired to the transmitter"
- "Grab-sample valve at the flow cell"
+constraint: all_of
```
…34 unchanged lines
```datasheet
−label: Minimum Reagent Replacement Interval
+label: Reagent Replacement Interval
type: range
unit: days
…2 unchanged lines
max: 180
setpoints: [14, 30, 60, 90, 180]
+constraint: min
```
…101 unchanged lines
max: 4
setpoints: [0, 2, 4]
+constraint: exact
```
…133 unchanged lines
- "Sample flow and low-sample-flow switch verified"
- "Temperature compensation verified"
+constraint: all_of
```
…18 unchanged lines
max: 8
setpoints: [1, 2, 4, 8]
+constraint: min
```
…25 unchanged lines
setpoints: [1, 2, 3, 5]
default: 1
+constraint: min
```
…27 unchanged lines
- "One replacement reagent set for each wet-chemistry analyzer"
- "Calibration buffers and reference standards for one year of calibrations"
+constraint: all_of
```
## Spare sensors shall be the same model as the installed sensors, shall be delivered in the manufacturer's storage packaging, and shall be labeled with the parameter and the tags of the installed sensors they replace.

View current revision